Validity of non-invasive point-of-care hemoglobin estimation in healthy and sick children-a method comparison study.

2016 
This study was conducted at a tertiary care center in northern India to evaluate the validity of non-invasive transcutaneous hemoglobin estimation in healthy and sick children in comparison to hemoglobin estimation by traditional lab method. A method comparison study was conducted including 150 subjects. Enrolled patients included 80 neonates with average age of 3.9 ± 2.1 days, and 70 children with average age of 5.8 ± 2 years. Each population (newborn and children) comprised of almost equal numbers of healthy and critically ill patients with shock. Hemoglobin (Hb) was estimated on enrolment by transcutaneous spectrophotometry (SpHb) and traditional automated lab analyzer (Hb-Lab). Difference between Hb levels by the two methods (called bias) was measured and analyzed using Bland-Altman method. Out of 148 data pairs analyzed, bias between SpHb and Hb-Lab was −1.52 ± 1.91 g/dl (mean ± SD). SpHb showed excellent positive correlation with Hb-Lab (r = 0.94 (p < 0.001)) and good visual agreement on Bland-Altman plots. Bias was higher in sick subjects with shock as compared to healthy ones in both neonatal and pediatric population (−2.31 ± 2.21 g/dl versus −0.77 ± 1.2 g/dl, respectively).