Spatial and contrast resolution of ultralow dose dentomaxillofacial CT imaging using iterative reconstruction technology

2017 
Objectives:The objective of this study was to determine how iterative reconstruction technology (IRT) influences contrast and spatial resolution in ultralow-dose dentomaxillofacial CT imaging.Methods:A polymethyl methacrylate phantom with various inserts was scanned using a reference protocol (RP) at CT dose index volume 36.56 mGy, a sinus protocol at 18.28 mGy and ultralow-dose protocols (LD) at 4.17 mGy, 2.36 mGy, 0.99 mGy and 0.53 mGy. All data sets were reconstructed using filtered back projection (FBP) and the following IRTs: adaptive statistical iterative reconstructions (ASIRs) (ASIR-50, ASIR-100) and model-based iterative reconstruction (MBIR). Inserts containing line-pair patterns and contrast detail patterns for three different materials were scored by three observers. Observer agreement was analyzed using Cohen's kappa and difference in performance between the protocols and reconstruction was analyzed with Dunn's test at α = 0.05.Results:Interobserver agreement was acceptable with a mean kappa ...