Validation of the Pangao PG-800B26 upper arm blood pressure monitor in the general population according to the European Society of Hypertension and the British Hypertension Society protocols

2017 
OBJECTIVE: This study aimed to evaluate the accuracy of the automated oscillometric upper arm blood pressure monitor Pangao PG-800B26 for home blood pressure monitoring according to the European Society of Hypertension International Protocol (ESH-IP) revision 2010 and the British Hypertension Society (BHS) protocols. PARTICIPANTS AND METHODS: Systolic and diastolic blood pressures were measured sequentially in 33 and 85 adults, respectively, on the basis of the ESH-IP and BHS protocols using a mercury sphygmomanometer (two observers) and the device (one supervisor). The procedures and analysis methods of the protocols were followed precisely. RESULTS: The device fulfilled the criteria of the ESH-IP, with device-observer differences of 1.01±5.16 and -0.58±4.17 mmHg for systolic and diastolic blood pressure, respectively. Furthermore, the A/A grade of the BHS protocol was also achieved for overall grading and for the three pressure levels, with average differences of 0.85±6.35 and -0.15±5.65 mmHg for systolic and diastolic blood pressure, respectively, which also fulfilled the requirements of the Association for the Advancement of Medical Instrumentation. CONCLUSION: The Pangao PG-800B26 fulfilled the criteria of the ESH-IP 2010 and achieved the A/A grade of the BHS protocol, and hence can be recommended for home use in adults.