Evaluation of a novel feline AB blood typing device

2014 
This prospective study evaluated a novel immunochromatographic (IC) blood typing test for the AB blood group system. Typing was conducted comparatively on ethylenediamine tetra-acetic acid-anticoagulated blood samples from 89 sick and 16 healthy cats with the IC test, as well as two tests as reference methods, a tube agglutination and a gel column test. The samples were between 0 and 10 days old (median 3 days) and were tested for haemolysis and agglutination; the packed cell volume ranged from 0.07 to 0.57 l/l (median 0.40 l/l). The reference methods agreed with each other in 100% of the test runs. Of the 85 samples tested as blood type A by the two reference methods, 80 were correctly identified by the IC test, four were misidentified as AB and one was rated inconclusive. All B samples were correctly typed. Two of the three AB samples were correctly identified by the IC test and one was rated inconclusive. The sample quality had no influence on test performance. Of 30 repeats, 28 were readable and showe...