Performance of blood glucose meters in compliance with current and future clinical ISO15197 accuracy criteria

2014 
AbstractSince 2003, blood glucose meters for patient self testing are approved in Europe based on the accuracy performance criteria as defined by the ISO15197 guideline. A new draft ISO guideline is currently under regulatory review, which suggests more strict accuracy acceptance criteria, and which may not be entirely fulfilled by currently commercialized blood glucose meter systems. In order to investigate the compliance of BG*Star and iBG*Star and several other blood glucose meters with the new draft ISO guideline, we performed a post-hoc analysis of data obtained from a recently performed ISO-conforming clinical accuracy performance study. This study was performed with 106 patients, clinically presenting with blood glucose levels distributed over the entire measurement range and in line with the glucose distribution requirements as demanded by the guideline. The YSI 2300 STAT Plus analyzer (glucose oxidase) served as reference method.While all tested meters had been in a high degree of compliance with...